Customization: | Available |
---|---|
Type: | Infrared Gas Sensor |
Conductivity Type: | Surface Type |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Technical Specifications | ||
MEASUREMENT | ||
Operating Principle | 3-electrode electrochemical | |
Detection Range | 0~5 ppm | |
Sensitivity | 2.0 ± 1.0 μA/ppm | |
Response Time (T90) | ≤45s | |
Repeatability | <±2% signal | |
Linearity | Linear | |
MEASUREMENT | <2% signal/month | |
ELECTRICAL |
||
Resolution | 0.02 ppm | |
Recommended Load | 5~35 | |
Bias Voltage | 0mV | |
ENVIRONMENTAL |
||
Operating Temp. Range | -20°C ~ 50°C | |
Operating Humidity Range | 15% RH ~ 90% RH non- condensing | |
Operating Pressure Range | 800 ~ 1200 mbar | |
LIFETIME |
||
Recommended Storage Temp. | 0°C to +20°C in sealed container | |
Expected Operating Life | 24 months in air | |
Storage Life | 6 months in original packaging | |
Standard Warranty | 12 months from date of despatch |
Cross-sensitivity Data |
||
Gas | Concentration(ppm) | Output signal (ppm SiH3Cl equivalent) |
Hydrogen | 100 | 0 |
Sulfur dioxide | 1 | 2 |
Methane | 1000 | 0 |
Carbon dioxide | 3000 | 0 |
Carbon monoxide | 50 | 0 |